UHPLC Analysis of A Monoclonal Antibody System Suitability Standard on BIOshell IgG 1000 Å C4

CONDITIONS
column
BIOshell IgG 1000 Å C4; 10 cm x 2.1 mm I.D., 2.7 μm particles (63288-U)
mobile phase
[A] 70:30 0.1% TFA (water): 0.1% TFA (acetonitrile); [B] 50:50 0.1% TFA (water): 0.1% TFA (acetonitrile)
gradient
0% B to 50% B in 50 min; 50% B to 100% B in 10 min
flow rate
0.2 mL/min
pressure
1325 psi (91.4 bar)
column temp.
75 °C
detector
UV, 215 nm
injection
1 μL
sample
MMMonoclonal Antibody System Suitability Standard, 1 g/L, 0.05% TFA (water)Monoclonal Antibody System Suitability Standard, 1 g/L, 0.05% TFA (water) A03
4
A034Monoclonal Antibody System Suitability Standard, 1 g/L, 0.05% TFA (water) A034MMMonoclonal Antibody System Suitability Standard, 1 g/L, 0.05% TFA (water)Monoclonal Antibody System Suitability Standard, 1 g/L, 0.05% TFA (water)
A03
说明
分析说明
This application demonstrates the ability of the BIOshell IgG 1000 Å C4 column to resolve different monoclonal antibody (mAb) variants. The mAb variants within this sample differ by their pattern of glycosylation.