GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) in Soybean Oil on SLB®-35ms after SPE using Supelclean EZ-POP NP

GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) in Soybean Oil on SLB®-35ms after SPE using Supelclean™ EZ-POP NP suitable for GC, application for SPE

材料

used together

产品编号
说明
价格

Supelclean EZ-POP NP小柱

volume 12 mL, pkg of 20

相关产品

产品编号
说明
价格

苯并[a]芘

≥96% (HPLC)

荧蒽

analytical standard

analytical standard

98%

苯并[k]荧蒽

analytical standard

analytical standard

analytical standard

苊烯

analytical standard

analytical standard

analytical standard, for environmental analysis

苯并[b]荧蒽

analytical standard

茚并[1,2,3-cd]比

analytical standard

进样口衬管,分流/不分流,具有锥部的 FAST FocusLiner(填充玻璃棉)

pkg of 1 ea

analytical standard

二苯并[a,h]蒽溶液

certified reference material, 100 μg/mL in methylene chloride

多环芳烃混标

certified reference material, 2000 μg/mL each component in benzene: dichloromethane (50:50), ampule of 1 mL

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

soybean oil spiked with PAHs at 10 ng/g

SPE tube/cartridge

Supelclean EZ-POP NP, 2.5 g/12 mL (54341-U)

condition

10 mL acetone, dry under 10-15" Hg for 10 minutes

sample addition

0.5 mL weighed accurately onto cartridge

elution

15 mL acetonitrile

eluate post-treatment

concentrate to final volume of 0.5 mL at 40 °C under nitrogen

column

SLB-35ms, 30 m x 0.25 mm I.D., 0.25 μm (29804-U)

oven

60 °C (1 min), 20 °C/min to 340 °C (10 min)

inj. temp.

300 °C

detector

MSD-SIM (MS source at 250 °C, quads at 200°C)

MSD interface

330 °C

scan range

SIM

carrier gas

helium, 1 mL/min constant flow

liner

2 mm I.D. FocusLiner with taper

说明

分析说明

PAHs are ubiquitous in the environment, resulting from both natural and manmade sources of combustion. They can contaminate food crops through environmental exposure to particulates, automobile exhaust, etc. In the case of edible oils, additional PAHs can be introduced during actual production processes such as seed drying. In this application, EZ-POP NP SPE is used to extract PAHs from two popular edible oils, canola and soybean. The 16 PAHs designated in many US EPA methods were analyzed by GC/MS after extraction from spiked samples of oil.