跳转至内容
Merck
CN

GC Analysis of Volatiles in White Wine on VOCOL® after SPME using a 75 μm Carboxen/PDMS Fiber

GC Analysis of Volatiles in White Wine on VOCOL® after SPME using a 75 μm Carboxen/PDMS Fiber suitable for GC, application for SPME

材料

used together

产品编号
说明
价格

Methyl formate

analytical standard

SPME Fiber Assembly

75 μm CAR/PDMS, Fused Silica (1 cm), needle size 24 ga, Manual Holder, pk of 3, black hub

相关产品

产品编号
说明
价格

Hexyl acetate

analytical standard

Octanoic acid

analytical standard

Ethyl butyrate

analytical standard

Isoamyl alcohol

≥98%, FG

2-Methyl-1-butanol

≥99%

Sulfur dioxide

≥99.9%

VOCOL® Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 1.50 μm

SPME Fiber Holder

for use with Manual Sampling, pk of 1

SPME Fiber Holder

for use with CTC CombiPAL, Gerstel MPS2 Autosamplers, pk of 1

Ethyl hexanoate

analytical standard

Ethyl octanoate

analytical standard

2,3-Butanediol

analytical standard, mixture of racemic and meso forms

CONDITIONS

sample/matrix

white wine + 25% NaCl

SPME fiber

Carboxen/PDMS, 75 μm (57318)

extraction

10 min, 40 °C (headspace)

desorption process

3 min at 290 °C

column

VOCOL, 30 m × 0.25 mm I.D., 1.5 μm (24205-U)

oven

40 °C (2 min) to 220 °C at 10 °C/min

inj. temp.

290 °C

detector

GC-MS, quadrupole, m/z = 31-240

carrier gas

helium, 35 cm/sec

injection

splitless/split, closed 1 min

liner

0.75 mm liner

sample

white wine

说明

分析说明

In this application, SPME was used for the analysis of volatiles in white wine using a headspace method. The presence of aroma compounds in white wine is one of the most important attributes of its final quality. A Carboxen/PDMS fiber was inserted into the headspace of the vial for 10 minutes at 40 °C and analyzed by GC/MS. The thick film of the VOCOL GC column helps focus the volatile analytes.

法律信息