HPLC Analysis of Tetramisole Enantiomers and Related Compounds on Astec® CYCLOBOND I 2000 DMP

HPLC Analysis of Tetramisole Enantiomers and Related Compounds on Astec® CYCLOBOND™ I 2000 DMP application for HPLC

材料

分析色谱柱

产品编号
说明
价格

Astec® CYCLOBOND I 2000 DMP 手性 HPLC 色谱柱

5 μm particle size, L × I.D. 25 cm × 4.6 mm

标准

产品编号
说明
价格

四咪唑 盐酸盐

phosphatase inhibitor

可卡因标准液 溶液

1.0 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®

(−)-Levamisole hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Astec CYCLOBOND I 2000 DMP 10 cm X 2.1 mm

mobile phase

(A) 100 mM ammonium acetate, pH 5 with acetic acid (B) acetonitrile, 90:10, v/v

flow rate

0.2 mL/min

pressure

360 psi

column temp.

35 °C

detector

UV, 220 nm

injection

2 μL

sample

100 ug/mL tetramisole and 50 ug/mL each cocaine and aminorex in water:methanol, 80:20, v/v

说明

分析说明

Cocaine is often found laced with levamisole (one of the tetramisole enantiomers) and aminorex is a metabolite of levamisole known to cause serious vascular damage in drug abusers. These compounds are readily separated using the Astec CYCLOBOND column - a derivatized cyclodextrin column only offered by Supelco. What is really interesting about this separation is that it shows the ability of chiral HPLC columns to separate enantiomers (here, dexamisole and levamisole) by chiral discrimination, as well as non-chiral discrimination in the same analysis. The compounds aminorex and cocaine are not enantiomers of each other or of tetramisole. The highest grade LC-MS solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.