Stand-alone, for use with Supelguard cartridges (2 cm L. x 2.1 to 4.0 mm I.D.)CONDITIONS
column
Ascentis RP-Amide, 25 cm x 4.6 mm I.D.,5 μm particles (565325-U)
mobile phase
A [A] DI water; [B] acetonitrile; (60:40, A:B)
gradient
0 min 75% A; 15 min 5% A; 0 min 25% B; 15 min 95% B; A:B held for 5 min
flow rate
1.5 mL/min
column temp.
40 °C
detector
UV, 360 nm
injection
10 μL
sample
(as listed in mobile phase)
mobile phase
B [A] DI water; [B] acetonitrile; (25:75, A:B)
说明
分析说明
This application demonstrates the suitability of the Ascentis RP-Amide for the analysis of 21 aldehyde/ketone DNPH derivatives.