跳转至内容
Merck
CN

US EPA Method 8315 (Solid Waste Carbonyls by HPLC-UV) on Ascentis® RP Amide

US EPA Method 8315 (Solid Waste Carbonyls by HPLC-UV) on Ascentis® RP Amide application for HPLC

材料

used together

产品编号
说明
价格

Ascentis® C18 Supelguard Guard Cartridge

5 μm particle size, L × I.D. 2 cm × 4 mm, pkg of 2 ea

Ascentis® RP-Amide HPLC Column

3 μm particle size, L × I.D. 15 cm × 4.6 mm

Ascentis® RP-Amide HPLC Column

5 μm particle size, L × I.D. 25 cm × 4.6 mm

Ascentis® C18 Supelguard Guard Cartridge

3 μm particle size, L × I.D. 2 cm × 4 mm, pkg of 2 ea

Ascentis® C8 Supelguard Guard Cartridge

5 μm particle size, L × I.D. 2 cm × 4 mm, pkg of 2 ea

相关产品

产品编号
说明
价格

Ascentis® C18 HPLC Column

3 μm particle size, L × I.D. 15 cm × 4.6 mm

Ascentis® C18 HPLC Column

5 μm particle size, L × I.D. 25 cm × 4.6 mm

Ascentis® RP-Amide Supelguard Guard Cartridge

5 μm particle size, L × I.D. 2 cm × 4 mm, pkg of 2 ea

Ascentis® C8 HPLC Column

3 μm particle size, L × I.D. 15 cm × 4.6 mm

Ascentis® C8 HPLC Column

5 μm particle size, L × I.D. 25 cm × 4.6 mm

Ascentis® Express C18 (2.7 μm) HPLC Column

L × I.D. 15 cm × 4.6 mm, pk of 1 ea

Ascentis® Express C8, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 15 cm × 4.6 mm

配件

产品编号
说明
价格

Wheaton wide-mouth bottle

amber soda-lime glass bottle, capacity 1000 mL, white polypropylene cap, (PTFE faced LDPE liner), case of 12 ea

Coupler for Legacy Guard Column Holder

PEEK, I.D. × O.D. × Overall L 0.010 in. × 1/16 in. × 1 in.

Supelguard Guard Cartridge Holder

Stand-alone, for use with Supelguard cartridges (2 cm L. x 2.1 to 4.0 mm I.D.)

CONDITIONS

column

Ascentis RP-Amide, 25 cm x 4.6 mm I.D.,5 μm particles (565325-U)

mobile phase

A [A] DI water; [B] acetonitrile; (60:40, A:B)

gradient

0 min 75% A; 15 min 5% A; 0 min 25% B; 15 min 95% B; A:B held for 5 min

flow rate

1.5 mL/min

column temp.

40 °C

detector

UV, 360 nm

injection

10 μL

sample

(as listed in mobile phase)

mobile phase

B [A] DI water; [B] acetonitrile; (25:75, A:B)

说明

分析说明

This application demonstrates the suitability of the Ascentis RP-Amide for the analysis of 21 aldehyde/ketone DNPH derivatives.

法律信息

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany