UHPLC/MS Analysis of Ten Cannabinoids on Ascentis® Express RP-Amide

材料
相关产品
(−)-Δ9-四氢大麻酚标准液 溶液
1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®次大麻二酚(CBDV)标准液 CRM 溶液
1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®大麻萜酚酸(CBGA)标准液 CRM 溶液
1.0 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®大麻二酚酸标准液 CRM 溶液
1.0 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®次四氢大麻酚标准液 CRM 溶液
1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®-四氢大麻酚酸标准液 溶液
1.0 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®CONDITIONS
column
Ascentis Express RP-Amide, 10 cm x 2.1 mm I.D., 2.0 μm particles (53913-U)
mobile phase
[A] 0.1% formic acid; [B] 0.1% formic acid in acetonitrile
gradient
60% B to 100% B in 10 min
flow rate
0.4 mL/min
pressure
7600 psi (524 bar)
column temp.
35 °C
detector
MS, ESI(+), ESI(-), MRM
injection
1 μL
sample
1000 ng/mL each in methanol
说明
分析说明
Cannabis compounds reportedly have therapeutic efficacy in the treatment of pain, mood disorders, and inflammatory diseases. These standards are used in testing methods by GC/MS, LC/MS, or HPLC for applications in clinical toxicology, testing of cannabis potency or impurity profiling by growers, pharmaceutical research, forensic analysis, and urine drug testing. Shown here is the separation of cannabis compounds on an Ascentis Express RP-Amide column. Highest grade UHPLC solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.