UHPLC/MS Analysis of Ten Cannabinoids on Ascentis® Express RP-Amide

UHPLC/MS Analysis of Ten Cannabinoids on Ascentis® Express RP-Amide application for UHPLC-MS

材料

相关产品

产品编号
说明
价格

Ascentis® Express RP-Amide 2.7μm HPLC 色谱柱

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

大麻二酚标准液 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

大麻酚标准液 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

(−)-Δ9-四氢大麻酚标准液 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

次大麻二酚(CBDV)标准液 CRM 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

大麻萜酚标准液 CRM 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

大麻萜酚酸(CBGA)标准液 CRM 溶液

1.0 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®

大麻色烯标准液 CRM 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

大麻二酚酸标准液 CRM 溶液

1.0 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®

次四氢大麻酚标准液 CRM 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

-四氢大麻酚酸标准液 溶液

1.0 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Ascentis Express RP-Amide, 10 cm x 2.1 mm I.D., 2.0 μm particles (53913-U)

mobile phase

[A] 0.1% formic acid; [B] 0.1% formic acid in acetonitrile

gradient

60% B to 100% B in 10 min

flow rate

0.4 mL/min

pressure

7600 psi (524 bar)

column temp.

35 °C

detector

MS, ESI(+), ESI(-), MRM

injection

1 μL

sample

1000 ng/mL each in methanol

说明

分析说明

Cannabis compounds reportedly have therapeutic efficacy in the treatment of pain, mood disorders, and inflammatory diseases. These standards are used in testing methods by GC/MS, LC/MS, or HPLC for applications in clinical toxicology, testing of cannabis potency or impurity profiling by growers, pharmaceutical research, forensic analysis, and urine drug testing. Shown here is the separation of cannabis compounds on an Ascentis Express RP-Amide column. Highest grade UHPLC solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.