跳转至内容
Merck
CN

LC/MS Analysis of Isoflavones on Chromolith® FastGradient RP-18 endcapped

LC/MS Analysis of Isoflavones on Chromolith® FastGradient RP-18 endcapped application for LC-MS

材料

相关产品

产品编号
说明
价格

Genistein

synthetic, ≥98% (HPLC), powder

Genistein

from Glycine max (soybean), ~98% (HPLC)

Puerarin

analytical standard

Daidzein

≥98%, synthetic

Puerarin

≥98.0% (HPLC)

Acetonitrile solution

contains 0.1 % (v/v) formic acid, suitable for HPLC

Glycitin

≥98% (HPLC)

Daidzein

analytical standard

Genistein

analytical standard

Puerarin

analytical standard

Glycitin

United States Pharmacopeia (USP) Reference Standard

Genistein

United States Pharmacopeia (USP) Reference Standard

Daidzein

United States Pharmacopeia (USP) Reference Standard

CONDITIONS

column

Chromolith FastGradient RP-18, 5 cm x 2.0 mm I.D., 2 μm particles (1.52007.0001)

mobile phase

[A] 0.1% formic acid in water; [B] 0.1% formic acid in acetonitrile

gradient

5 to 15% B in 1.5 min, to 95% B in 4 min, held at 95% B for 1 min, to 5% B in 1 min, held at 5% for 1.5 min

flow rate

0.5 mL/min

pressure

821-317 psi (57 - 22 bar)

column temp.

25 °C

detector

ESI-MS(+) (m/z range 100-600), overlay of four EICs (m/z 271.0, 255.0, 447.1, 417.1)

injection

1 μL

sample

genistein 10 ppb, daidzein 10 ppb, glycitin 10 ppb, puerarin 20 ppb in methanol and water (50:50 v/v)

说明

法律信息